# Example: Dose scoring using the Track-Length Estimator #
# TG186 source inside the TG186 applicator #
# Water cube of 50 cm side #
# Centered dose grid #
# #
# #
# Latest version - May 2021 #
# Author: Francisco Berumen #
# Contact - fberumenm@gmail.com #
# https://doi.org/10.1016/j.brachy.2020.12.007 #
# #
##### Import the source inside the applicator #####
includeFile = HDRSourceInApplicator.txt
##### Scorer geometry ######
s:Ge/DoseGrid/Type = "TsBox"
s:Ge/DoseGrid/Parent = "World"
d:Ge/DoseGrid/HLX = 10.05 cm
d:Ge/DoseGrid/HLY = 10.05 cm
d:Ge/DoseGrid/HLZ = 10.05 cm
s:Ge/DoseGrid/Color = "red"
b:Ge/DoseGrid/IsParallel = "T"
i:Ge/DoseGrid/XBins = 201
i:Ge/DoseGrid/YBins = 201
i:Ge/DoseGrid/ZBins = 201
##### Dose scoring using the TLE ####
s:Sc/DoseOnRTGrid/Quantity = "TrackLengthEstimator"
sv:Sc/DoseOnRTGrid/Report = 1 "Sum"
s:Sc/DoseOnRTGrid/InputFile = "Muen.dat"
s:Sc/DoseOnRTGrid/Component = "DoseGrid"
b:Sc/DoseOnRTGrid/OutputToConsole = "F"
s:Sc/DoseOnRTGrid/IfOutputFileAlreadyExists = "Increment"
s:Sc/DoseOnRTGrid/OutputType = "DICOM"
b:Sc/DoseOnRTGrid/DICOMOutput32BitsPerPixel = "True"
s:Sc/DoseOnRTGrid/OutputFile = "DoseTLE"
sv:Sc/DoseOnRTGrid/OnlyIncludeParticlesNamed = 1 "gamma"
##### Local energy deposition of electrons
d:Ph/Default/CutForElectron = 10 cm
##### Run 10000 histories
b:Gr/Enable = "F"
i:So/Source/NumberOfHistoriesInRun = 10000